Repairing the Sampler
4
1100 Series WS MWS Reference Manual
139
Injection-Valve Assembly
Removing the injection valve assembly
1
In the user interface start the maintenance mode and select the “Change
Needle/Seat” function.
2
Switch off the sampler at the main power switch. Remove the power cable.
3
Remove the top cover, top plate, and foam (see
4
Remove the plate tray from the tray base.
5
Remove the transport assembly, (see
“Removing the sample transport"
6
Remove the sampling unit, (see
7
Disconnect the valve-cable connector from the SUD board (Valve Assy).
8
Remove the three screws which fix the injection valve assembly.
9
Remove the injection valve assembly.
When required
When defective
Tools required
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
Parts required
0101-0921 Injection Valve assembly
for the G1367A/68A samplers
0101-1050 Micro Injection valve assembly
for the G1377A/78A samplers
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......