Repairing the Sampler
4
1100 Series WS MWS Reference Manual
155
Replacing the Autosampler Firmware
The installation of new firmware is required:
•
if new version solves problems of currently installed version.
•
if after exchange of the mainboard (MTP) the version on board is older than
previous installed one.
To upgrade the autosampler firmware the following steps have to be
performed:
1
Load the firmware into the autosampler, see the help system of your user
interface.
For control module see
2
If the sampler main board was replaced, re-enter the serial number
information of the module trough the user interface, see
Serial Number using the Control Module"
on page 148 or