Introduction to the Well-plate Sampler
6
1100 Series WS MWS Reference Manual
201
The standard sampling sequence occurs in the following order:
1
The injection valve switches to the bypass position.
2
The plunger of the metering device moves to the initialization position.
3
The needle lock moves up.
4
The needle moves to the desired sample vial position.
5
The needle lowers into the vial.
6
The metering device draws the preset sample volume.
7
The needle lifts out of the vial.
8
The needle is then moved onto the seat to close the sample loop.
9
The needle lock moves down.
10
The injection cycle is completed when the injection valve switches to the
mainpass position.
When using an injector program the lines 3 to 6 are replaced by the contents of
the injector program.
If needle wash is required it will be done between step 6 and 7.