Theory of Operation
7
1100 Series WS MWS Reference Manual
229
Analog Signal Output
The analog signal output can be distributed to a recording device. For details
refer to the description of the module’s main board.
GPIB Interface
The GPIB connector is used to connect the module with a computer. The
address and control switches next to the GPIB connector determine the
GPIB address of your module. The switches are preset to a default address and
recognized by the operating software from Agilent Technologies.
CAN Interface
The CAN is an intermodule communication interface. It is a 2-wire serial bus
system supporting high speed data communication and real-time requirement.
N O T E
A chemstation cannot be connected directly to the well-plate sampler by GPIB.
Table 66
Default Addresses
Autosampler
28
RID
29
Pump
22
FLD
23
VWD
24
Autosampler (HP 1050)
18
Agilent 8453A
25
Pump (HP 1050)
16
DAD / MWD
26
VWD (HP 1050)
10
Column Compartment
27
DAD (HP 1050)
17
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......