Troubleshooting and Test Functions
3
1100 Series WS MWS Reference Manual
83
Motor Temperature
One of the motors of the transport assembly has drawn excessive current,
causing the motor to become too hot. The processor has switched off the
motor to prevent damage to the motor.
See figure for motor identification.
Motor 0 temperature
: X-axis motor.
Motor 1 temperature
: Z-axis motor.
Motor 2 temperature
: Theta (gripper rotation) motor.
The processor monitors the current drawn by each motor and the time the
motor is drawing current. The current drawn by the motors is dependent on
the load on each motor (friction, mass of components etc.). If the current
drawn is too high, or the time the motor draws current is too long, the error
message is generated.
Probable Causes
•
Mechanical obstruction.
•
High friction in the transport assembly.
•
Motor belt tension too high.
•
Defective motor.
•
Defective transport assembly flex board.
Suggested Actions
✔
Switch off the sampler at the power switch. Wait at least 10 minutes before
switching on again.
✔
Ensure unobstructed movement of the transport assembly.
✔
Exchange the sample transport assembly.
✔
Exchange the MTP main board.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......