Troubleshooting and Test Functions
3
1100 Series WS MWS Reference Manual
89
Sample Transport Self Alignment
The sample transport alignment with the sampling unit and the well-plate tray
is required to compensate for larger deviations in positioning the needle
carrier.This function is in the diagnose screen of the Chemstation or the
Control Module.
The sample transport self alignment is required after disassembling the
system or when you exchange:
•
The sample transport.
•
The sampling unit.
•
The MTP main board.
•
The well plate sampler tray
Table 15
Maintenance positions
Function
Arm position in X
Arm position in
Theta
Arm Position in Z
Note
Change Needle
Left side
Straight
Up
No current on Theta
Change Carrier assembly
Left side
Straight
Middle
No current on the ST
Change Loop capillary
Middle
Left
Up
Home position
Right side
Left rear
Up
Park arm
Left side
Right rear
Up
WA R N I N G
The sample transport self alignment requires the standard well-plate tray (Part
Number: G1367-60001)
WA R N I N G
The sample transport self alignment must be done with an empty WPS tray
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......