Troubleshooting and Test Functions
3
1100 Series WS MWS Reference Manual
95
Errors which may occur during the turn ON and initialization process
Step 1
Symptom
“Failure to turn ON”
No activity when power button is pushed on. Power indicator stays off.
Possible causes
•
Defective main board
•
Power supply defective
Suggested actions
✔
Turn the WPS off. Disconnect the power supply from the main board, see if
power indicator becomes green when the WPS is turned-on.
•
If yes, change the main board (G1367-69500)
•
If no, change the power supply (0905-2528 Rev G or higher)
Step 2
Symptom
“Fan failure”
Early during the turn-on/initialization process a main fan or an exhaust fan
error occurs.
Possible causes
•
Fan not connected to the main board
•
Fan connector defective
•
Fan defective
•
Main board defective