98
1100 Series WS MWS Reference Manual
3
Troubleshooting and Test Functions
Step 5
Symptom
”Initialization fails”
The initialization fails to complete its required movements, resulting in one or
more of a variety of possible error messages. The error message produced
depends on when the failure occurred during the initialization.
Gather the revisions information, user interface error and logbook
information, and error code information as described on page 1. Also,
consider anything which may have been done just before the initialization
failure occurred.
To date, we have seen initialization errors occur for a variety of reasons. As an
overall troubleshooting plan, follow the suggested actions described bellow.
If this does not help, take the information from the instrument logbook, refer
to the appropriate section in the next chapter and follow the recommended
actions step by step.
Suggested actions
✔
Turn the WPS off, manually change the X-position of the sample transport
and rotate the needle carrier to a different position, turn-on the WPS again.
✔
If there is any weight on top of the WPS, remove that weight, turn-off the
WPS and then turn it on again.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......