Parts and Materials
5
1100 Series WS MWS Reference Manual
185
Agilent 1100 to 3396 Series II / 3395A Integrators
Use the cable
03394-60600
and cut pin #5 on the integrator side. Otherwise
the integrator prints START; not ready.
N O T E
START and STOP are connected via diodes to pin 3 of the 3394 connector.
Table 44
Agilent 1100 to 3396A Integrators
Connector
03394-60600
Pin
3394
Pin
Agilent 1100
Signal Name
Active
(TTL)
9
1 - White
Digital ground
NC
2 - Brown
Prepare run
Low
3
3 - Gray
Start
Low
NC
4 - Blue
Shut down
Low
NC
5 - Pink
Not connected
NC
6 - Yellow
Power on
High
5,14
7 - Red
Ready
High
1
8 - Green
Stop
Low
NC
9 - Black
Start request
Low
13, 15
Not connected