1100 Series WS MWS Reference Manual
1
Installing the Sampler
Installing a Thermostatted Sampler
1
Place the thermostat on the bench.
2
Remove the front cover and route the condensation drain tube to the waste
bottle.
3
Install the LAN interface board in the sampler (if required), see
4
Remove the adhesive tape which covers the side and front doors.
5
Open the front door and remove the left side door.
6
Remove the transport protection foam.
7
Re-install the corrugated waste tube in the plastic port.
Preparation
Locate bench space
Provide power connections
Unpack the sampler and the thermostat
Parts required
Sampler and thermostat
Power cord, for the other cables see below and,
ChemStation and/or Control Module G1323B
WA R N I N G
Make sure that the condensation drain tube is always above the liquid level in the
vessel. If the tube is located in liquid the condensed water cannot flow out of the
tube and the outlet is blocked. Any further condensation will then remain in the
instrument. This may damage the instruments electronics.
Figure 6
Condensation leak outlet
Condensation drain tube
Waste bottle
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......