background image

Agilent Technologies

Installing a G2916A Sampler Tray on a 
5890 Series II Gas Chromatograph

G2916A Kit

This procedure assumes that the Gas Chromatograph (GC) is either

A 5890A GC which has been fully upgraded to a 5890 Series II instrument 
(“fully” means that both electronic and mechanical upgrades have been 
performed; a printed circuit board-only upgrade does not qualify).

• OR

 a 5890 Series II GC which was purchased as such.

and that the G2916-61000 injector bracket has already been installed.

Summary of Contents for G2916A

Page 1: ...matograph GC is either A 5890A GC which has been fully upgraded to a 5890 Series II instrument fully means that both electronic and mechanical upgrades have been performed a printed circuit board only upgrade does not qualify OR a 5890 Series II GC which was purchased as such and that the G2916 61000 injector bracket has already been installed ...

Page 2: ...t side panel 2 Remove the pneumatic compartment access door from the left side panel 3 Reinstall the left side panel Table 1 Parts supplied Part number Description Quantity 18596 40015 Tray quadrants pkg of 4 1 G2614 60610 Cable main tray 1 G2916 60000 5890 Tray module 1 0100 1501 Tee inch hose for quadrant temperature control 2 G2916 80510 Tray screw kit Contains 1 Screw with washer M4 20 mm 4 51...

Page 3: ...t so that the frontmost pin on the injector bracket fits into the mating hole in the bottom of the tray then align the tray to the second alignment pin on the injector bracket CAUTION Do not move or manipulate the robotic arm or gripper Moving the arm or gripper back and forth can cause damage If necessary rotate the arm assembly a few degrees until there is enough clearance ...

Page 4: ...en the screws to secure the tray to the bracket 4 Install the tray quadrants Figure 2 Be sure to match the numbers on the quadrants with the tray base numbers Figure 1 Installing the tray Tray post Set tray on this alignment pin first Second alignment pin ...

Page 5: ...y then snap the front tab into the slot on the tray 6 Connect the tray cable from the back of the tray near the drain fitting to the Tray connector on the back of the G2912A Controller Figure 2 Install the tray quadrants Tray quadrant Front tab Guide Slot Tray base numbers ...

Page 6: ... United States and international copyright laws G2916 90000 First edition July 2004 Printed in USA Agilent Technologies Inc 2850 Centerville Road Wilmington DE 19808 1610 USA CAUTION A CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in damage to the prod uct or loss of important data Do not...

Reviews: