For Agilent DS 402 and DS 602 pumps
, align the front
edge of the vibration foam with the front end of the
pump base (Figure 5). Be sure that the sides of the pump
base align with the sides of the vibration foam.
4
Check your pump orientation inside the Quiet Cover DS.
The oil level window (sight glass) must face the front
door of the Quiet Cover DS. Make sure the pump power
cord is not crimped or under the pump.
Figure 5
Proper installation of the DS 402/602 pumps
Vibration foam
Pump base front edge
front edge