Agilent ADM Flow Meter Operating Instructions
31
Product Specifications
Other features
•
Automatic power off
•
Split ratio mode with continuous split flow rate reading
•
NIST Traceable Calibration
•
Compatible with dry gases
•
Volumetric and mass flow measurements
•
USB communication
•
Replaceable flow meter cartridge - individually computer calibrated
Table 2 ADM Flow Meter specifications
Flow range
0 to 750 mL/min, auto-ranging
Accuracy
*
:
* When verified at ambient temperature of 20 °C to 24 °C.
0 - 500 mL/min:
± 2% of reading, or ± 0.2 mL/min, whichever is
greater.
501 - 750 mL/min:
± 3% of reading.
Tubing temperature range:
-62 °C to 110 °C
Operating temperature range
0 °C to 45 °C
Storage Temperature range
†
:
† Damage to the instrument may occur if it is exposed to temperatures greater than
50 °C, leading to inaccurate results.
-15 °C to 50 °C
Power:
(3) AA batteries (alkaline), or USB power
Display:
128 x 64 Pixels Monochrome with 16 Gray
Scales