102
5977 Series MSD Operation Manual
4
Operating in Chemical Ionization (CI) Mode
3
Enter the reagent gas flow setpoint in the
Flow
field. This value is entered
as a percentage of maximum flow rate. The recommended flow is 20% for a
PCI source and 40% for an NCI source.
The flow control hardware remembers the flow setting for each gas. When
either gas is selected, the control board automatically sets the same flow
that was used for that gas the last time.
4
To begin reagent gas flow, select
Shutoff Valve
.
The system turns off the present gas flow while leaving the shutoff valve
(
103) open. This is to remove any residual gas in the
lines. Typical evacuation time is 6 minutes and then the shutoff valve is
closed.
The flow control module
The CI reagent gas flow control module regulates the flow of reagent gas into
the CI GC/MSD interface. The flow module consists of a mass flow controller
(MFC), gas select valves, CI calibration valve, shutoff valve, control electronics,
and plumbing. See
and
.
The back panel provides Swagelok inlet fittings for methane (
CH4
) and one
OTHER
reagent gas. The software refers to them as
Gas A
and
Gas B,
respectively. If you are not using a second reagent gas, cap the
OTHER
fitting to
prevent accidental admission of air to the analyzer. Supply reagent gases at
25
to
30
psi (170 to 205 kPa).
The shutoff valve prevents contamination of the flow control module by
atmosphere while the MSD is vented or by PFTBA during EI operation.
Summary of Contents for G7035A
Page 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Page 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Page 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Page 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Page 171: ......