Operating in Chemical Ionization (CI) Mode
4
5977 Series MSD Operation Manual
109
To Switch from the CI Ion Source to the Standard or Inert EI Ion Source
Procedure
1
From the Tune and Vacuum Control view, vent the MSD. See
76. The software will prompt you for the appropriate
actions.
2
Open the analyzer. See
“To Open the Analyzer Chamber”
3
Remove the CI/Extractor interface tip seal. See
4
Install the EI ion source. See
“To Install the EI Ion Source”
5
Place the CI ion source and interface tip seal in the ion source storage box.
6
Pump down the MSD. See
“To Pump Down the MSD in EI Mode”
7
Load your EI tune file.
C A U T I O N
Always wear clean gloves while touching the analyzer or any other parts that go inside
the analyzer chamber.
C A U T I O N
Electrostatic discharges to analyzer components are conducted to the side board
where they can damage sensitive components. Wear a grounded antistatic wrist strap
and take other antistatic precautions
before
you open the analyzer chamber.