Operating in Chemical Ionization (CI) Mode
4
5977 Series MSD Operation Manual
115
To Verify PCI Performance
Materials needed
•
Benzophenone, 100 pg/
L (8500-5440)
Procedure
1
Verify that the MSD performs correctly in E1 mode.
2
Verify that the
PCICH4.U
tune file is loaded.
3
Select
Gas A
and set flow to 20%.
4
In
Tune and Vacuum Control
view, perform CI setup. See
CI Ion Source to the Extractor EI Ion Source”
5
Run CI Autotune. See
6
Run the PCI sensitivity method
BENZ_PCI.M
using 1 µL of 100 pg/µL
benzophenone.
7
Verify that the system conforms to the published sensitivity specification.
Please see the Agilent Web site at
for
specifications.
C A U T I O N
Always verify MSD performance in EI before switching to CI operation. See
71. Always set up the CI MSD in PCI first, even if you are
going to run NCI.