5977 Series MSD Operation Manual
5
General Maintenance
To Remove the EI Ion Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Pliers, long-nose (8710-1094)
Procedure
1
Vent the MSD. See
2
Open the analyzer chamber. See
“To Open the Analyzer Chamber”
Make sure you use an antistatic wrist strap and take other antistatic
precautions before touching analyzer components.
3
If you are using a standard or inert EI ion source, disconnect the seven
wires from the ion source. Do not bend the wires any more than necessary
(See
Table 22
Standard EI ion source wires
Wire color
Connects to
Number of leads
Blue
Entrance lens
1
Orange
Ion focus
1
White
Filament 1 (top filament)
2
Red
Repeller
1
Black
Filament 2 (bottom filament)
2
C A U T I O N
Pull on the connectors, not on the wires.