General Maintenance
5
5977 Series MSD Operation Manual
145
To Assemble the Extractor EI Ion Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
•
Wrench, open-end, 10 mm (8710-2353)
Procedure
1
Slide the ceramic washer into the source body.
2
Insert the extractor lens into the source body, flat side first (See
3
Insert the entrance lens and ion focus lens into the insulator in the order
shown (
4
Slide the insulator containing the ion focus and entrance lens into the
source body, with the ion focus lens against the extractor lens (See
5
Install the two gold plated setscrews that holds the lenses in place.
6
Assemble the repeller assembly.
a
Install the repeller block insert into the source heater block assembly.
(See
b
Install the repeller insulators into the source heater block assembly and
repeller block insert.
c
Install the repeller through the repeller insulators, then put the flat
washer followed by the belleville spring washer onto the end of the
repeller shaft and secure finger tight with the repeller nut.
C A U T I O N
Do not overtighten the repeller nut or the ceramic repeller insulators will break when
the source heats up. The nut should only be finger tight.
Summary of Contents for G7035A
Page 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Page 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Page 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Page 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Page 171: ......