CI Maintenance
6
5977 Series MSD Operation Manual
169
5
Secure the new filament with the gold plated screw and washer.
6
After installing the filament, verify that it is not grounded to source body.
7
Install the ion source. See
“To Install the CI Ion Source”
8
Close the analyzer chamber. See
“To Close the Analyzer Chamber”
9
Pump down the MSD. See
“To Pump Down the MSD in CI Mode”
10
Perform a PCI autotune with methane. See
11
Select
Save Tune Parameters
from the
File
menu.