Installing GC Columns
2
5977 Series MSD Operation Manual
41
To Prepare the Column Ends for a CFT Fitting
Materials needed
•
SilTite ferrules:
•
0.1- to 0.25-mm columns, pkg of 10 (5188-5361)
•
0.32-mm columns, pkg of 10 (5188-5362)
•
0.53-mm columns, pkg of 10 (5188-5363)
•
Swaging nut
•
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
Procedure
1
Obtain the appropriate metal ferrule.
2
Pass the tubing end through the internal nut and SilTite ferrule leaving
approximately 1 cm of fused silica tubing protruding beyond the ferrule.
Thread the ferrule pre-swaging tool onto the internal nut with the tube
protruding (see
Figure 5
Assemble the internal nut, ferrule, and swaging nut
C A U T I O N
The SilTite ferrules are delicate. Follow the instructions in the next step very carefully
to avoid overtightening.