Operating in Electron Ionization (EI) Mode
3
5977 Series MSD Operation Manual
85
5
Press lightly on the side board to ensure a correct seal. Press on the metal
box on the side board.
The foreline pump will make a gurgling noise. This noise should stop within a
minute. If the noise continues, there is a
large
air leak in your system,
probably at the side plate seal, the interface column nut, or the vent valve.
6
Start the MassHunter Data Acquisition program.
7
In the
Instrument Control
view, from the Instrument menu, select
MS Vacuum
Control
to display the
Vacuum Control
dialog.
8
In the
Instrument Control
view
, Instrument
menu,
select
MS
Vacuum Control
to
display the
Vacuum Control
dialog.
9
Click
Pump Down
in the
Vacuum Control
dialog and follow the system prompts.
10
When prompted, turn on the GC/MSD interface heater and GC oven. Click
OK
when you have done so.
The software will turn on the ion source and mass filter (quad) heaters.
The temperature setpoints are stored in the current autotune (*.u) file.
11
After the message
Okay to run
appears, wait 2 hours for the MSD to reach
thermal equilibrium. Data acquired before the MSD has reached thermal
equilibrium may not be reproducible.
C A U T I O N
Do not turn on any GC heated zones until carrier gas flow is on. Heating a column with
no carrier gas flow will damage the column.
Summary of Contents for G7035A
Page 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Page 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Page 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Page 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Page 171: ......