3
Operating in Electron Ionization (EI) Mode
To Move or Store the MSD
Materials needed
•
Ferrule, blank (5181-3308)
•
Interface column nut (05988-20066)
•
Wrench, open-end, 1/4-inch × 5/16-inch (8710-0510)
Procedure
1
Vent the MSD (See
2
Remove the column and install a blank ferrule and interface nut.
3
Tighten the vent valve.
4
Move the MSD away from the GC (see the 5977 Series MSD
Troubleshooting and Maintenance Manual).
5
Unplug the GC/MSD interface heater cable from the GC.
6
Open the analyzer cover (See
7
Finger-tighten the side plate thumbscrews.
C A U T I O N
Do not overtighten the side plate thumbscrews. Overtightening will strip the threads in
the analyzer chamber. It will also warp the side plate and cause leaks.
Front thumbscrew
Rear thumbscrew