Operating in Chemical Ionization (CI) Mode
4
5977 Series MSD Operation Manual
97
To Switch from the Extractor EI Ion Source to the CI Ion Source
Procedure
1
Vent the MSD. See
2
Open the analyzer. See
“To Open the Analyzer Chamber”
3
Remove the extractor EI ion source. See
4
Remove the brown extractor wire from the feedthrough board and store it
with the EI extractor ion source. See
5
Install the CI ion source. See
“To Install the CI Ion Source”
6
Close the analyzer. See
“To Close the Analyzer Chamber”
7
Pump down the MSD. See
“To Pump Down the MSD in CI Mode”
C A U T I O N
Always verify MSD performance in EI before switching to CI operation.
Always set up the CI MSD in PCI first, even if you are going to run NCI.
C A U T I O N
Electrostatic discharges to analyzer components are conducted to the side board
where they can damage sensitive components. Wear a grounded antistatic wrist strap.
See
123. Take antistatic precautions
before
you
open the analyzer chamber.