General Maintenance
5
5977 Series MSD Operation Manual
137
To Clean the EI Ion Source
Materials needed
•
Abrasive paper (5061-5896)
•
Alumina abrasive powder (8660-0791)
•
Aluminum foil, clean
•
Cloths, clean (05980-60051)
•
Cotton swabs (5080-5400)
•
Glass beakers, 500 mL
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Solvents
•
Acetone, reagent grade
•
Methanol, reagent grade
•
Methylene chloride, reagent grade
•
Ultrasonic bath
Preparation
1
Disassemble the ion source. See
“To Disassemble the Standard or Inert EI
“To Disassemble the Extractor EI Ion Source”
2
Collect the following parts to be cleaned for a standard or inert EI ion
source: (See
•
Repeller
•
Interface socket
•
Source body
•
Drawout plate
•
Drawout cylinder
•
Ion focus lens
•
Entrance lens