General Maintenance
To Install the EI Ion Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Pliers, long-nose (8710-1094)
Procedure
1
Slide the ion source into the source radiator.
2
Install and hand tighten the source thumbscrews. Do not overtighten the
thumbscrews.
3
Connect the ion source wires as shown in
4
Close the analyzer chamber. See
“To Close the Analyzer Chamber”
Figure 27
Installing the EI ion source
Source feedthrough board
Ion source
Thumbscrews
Source heater and temperature
sensor wires
Source radiator
Extractor wire