Operating in Electron Ionization (EI) Mode
3
5977 Series MSD Operation Manual
71
To Verify System Performance
Materials needed
•
1 pg/µL (0.001 ppm) OFN sample (5188-5348)
Verify the tune performance
1
Verify that the system has been pumping down for at least 60
minutes.
2
Set the GC oven temperature to 150
°C and the column flow to 1.0
mL/min.
3
In the
Instrument Control
view, select
Checkout Tune
from the Checkout menu.
The software will perform an autotune and print the report.
4
When the autotune has completed, save the method and then select
Evaluate
Tune
from the
Checkout
menu.
The software will evaluate the last autotune and print a System
Verification – Tune report.
Verify the sensitivity performance
1
Set up to inject 1
µL of OFN, either with the ALS or manually.
2
In the
Instrument Control
view, select
Sensitivity Check
from the
Checkout
menu. The system displays an
Alert
dialog reminder about resolving the
OFN_SN method and placing the OFN sample in vial 1 when an ALS is
configured.
3
If necessary resolve your hardware with this method and place the sample
in the vial 1 position.
4
Click
OK
to run the method.
When the method is completed, an evaluation report will be printed.
Verify that rms signal-to-noise ratio meets the published specification. Please
see the Agilent Web site at www.agilent.com/chem for specifications.
Summary of Contents for G7035A
Page 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Page 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Page 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Page 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Page 171: ......