136
HP-IB Programming
OLA Specific Commands
parameter
1|2
1
Reads head A
2
Reads head B
Default is 1.
response
The measured value in dB or percent, depending on the current
setting of the measurement unit (see “:SENSe:POWer:UNIT”).
errors
The message “Wrong application for this command” is returned, if
the present application is not Coupler Test or Polarization
Dependent Coupler Test.
The message “No head connected” is returned, if there is only one
or no optical head attached to the instrument.
The message “No valid result possible” is returned, if the value
could not be measured.
example
OUTPUT 724;":SENS1:DATA? SR"
ENTER 724;A$
command
:SENSe:DATA? STAB
syntax
:SENSe[1|2]:DATA?<wsp>STAB
description
This query returns the stability value. It is only applicable to the
Stability application, which can only be started if an application is
active.
parameter
1|2
1
Reads head A
2
Reads head B
Default is 1.
response
The measured value in dB.
errors
The message “Wrong application for this command” is returned, if
the present application is not Stability.
The message “No head connected” is returned, if there is no optical
head attached to the specified channel.
The message “No valid result possible” is returned, if the value
cannot be measured.