HP-IB Programming
OLA Specific Commands
command
SENSe:POWer:REFerence:DISPlay:PPAR
syntax
SENSe[1|2]:POWer:REFerence:DISPlay:PPAR
description
This command stores the value currently measured and displayed as
the power of parasitic reflections. It is only applicable to the Return
Loss application.
parameter
1|2
1
Reads head A
2
Reads head B
Default is 1.
errors
The message “Wrong application for this command” is returned, if
the present application is not Return Loss.
The message “No head connected” is returned, if there is no optical
head attached to the specified channel.
related commands
:SENSe:POWer:REFerence:DISPlay:PPAR?,
:SENSe:POWer:REFerence,
:SENSe:POWer:REFerence:DISPlay:PREF
example
OUTPUT 724;"SENS1:POW:REF:DISP:PPAR"
command
:SENSe:POWer:REFerence:DISPlay:PPAR?
syntax
:SENSe[1|2]:POWer:REFerence:DISPlay:PPAR?
description
This query returns the stored value of the parasitic reflections. It is
only applicable to the Return Loss application.
parameter
1|2
The channel parameter 1 or 2 may be present, but is ignored. P
par is valid for both channels.
response
The stored parasitic reflections value P par.
related commands
:SENSe:POWer:REFerence:DISPlay:PPAR
errors
The message “Wrong application for this command” is returned, if
the present application is not Return Loss.
example
OUTPUT 724;":SENS:POW:REF:DISP:PPAR?"
ENTER 724;A$