Appendix C. Specifications
Technical Data, Product Specifications and Characteristics
Supplementary Performance Characteristics
Return Loss
Can be measured with internal source up to typically
55 dB with
±
1 dB accuracy.
General
Recalibration period: 2 years. Warm-up time to meet
accuracy specifications: 60 min. for instrument, 5 min with laser on.
HP-IB Capability
All modes and parameters can be
programmed, SCPI command set, HP 8153A lightwave multimeter
compliant command set where applicable.
HP-IB Interface Function Code
SH1, AH1, T6, L4, SR1, RL1,
PP0, DC1, DT0, C0.
Environmental
Storage temperature: –40
°
C to +70
°
C. Operating
temperature: 10
°
C to +40
°
C (35
°
C for HP81525A). Humidity:
< 95% R.H. within operating temperature range.
Power
100 – 240 Vrms, 100 VA max., 50 – 60 Hz.
Battery Back-up (for non-volatile memory)
With instrument
switched off all current modes and data will be maintained for at
least 10 years after delivery.
Dimensions
(WxHxD in cm): 42.6
×
8.9
×
44.5
Weight
9 kg