238
Appendix E. Cleaning Procedure
Cleaning Procedures
Cleaning Fragile Optical Devices
Some optical devices, such as the HP 81000BR Reference
Reflector, which has a gold plated surface, are very sensitive to
mechanical stress or pressure. Do not use cotton-swabs, soft-tissues
or other mechanical cleaning tools, as these can scratch or destroy
the surface.
Preferred Procedure
1
Blow away any dust or dirt with compressed air.
Procedure for Stubborn Dirt
To clean devices that are extremely sensitive to mechanical stress or
pressure you can also use an optical clean polymer film. This
procedure is time-consuming, but you avoid scratching or
destroying the surface.
1
Put the film on the surface and wait at least 30 minutes to make
sure that the film has had enough time to dry.
2
Remove the film and any dirt with special adhesive tapes.
Alternative Procedure
For these types of optical devices you can often use an ultrasonic
bath with isopropyl alcohol. Only use the ultrasonic bath if you are
sure that it won't cause any damage anything to the device.
1
Put the device into the bath for at least three minutes.
2
Blow away any remaining liquid with compressed air.
If there are any streaks or drying stains on the surface, repeat the
cleaning procedure.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com