Taking Standard Loss Measurements
Measuring the Insertion Loss
Insertion Loss Measurement Setup
The same setup can be
used to measure the
Polarization Dependent
Loss.
1
Connect the optical output to the input of the device under test
(DUT).
2
Connect the optical head which was used to measure and store
the reference power to the output of the DUT.
3
Activate the optical output.
If you wish to use the internal light source, press S
OURCE
O
N
/
O
FF
to turn the source on. The source indicator lamp must be lit.
If you have connected an external light source to the optical
input, press S
OURCE
O
N
/O
FF
to turn the internal source off. The
source indicator lamp must be off.
Figure 3-1
Setup for Insertion Loss Measurements
Head Input A
Head Input B
Optical Input
Optical Output
Head Input A
Head Input B
Optical Input
Optical Output
3 dB
Coupler
Pol. Ctrl.
Internal
Source(s)
Signal Processing and Display
Device
under
Test