110
6000 Series Oscilloscope User’s Guide
2
Front-Panel Controls
To restore the oscilloscope to its default configuration
•
Press the
Save/Recall
key, then press the
Default Setup
softkey.
The default configuration returns the oscilloscope to its default
settings. This places the oscilloscope in a known operating
condition. The major default settings are:
Horizontal
main mode, 100
µ
s/div scale, 0 s delay, center time
reference.
Vertical (Analog)
Channel 1 on, 5 V/div scale, DC coupling, 0 V
position, 1 M
Ω
impedance, probe factor to 1.0 if an AutoProbe
probe is not connected to the channel.
Trigger
Edge trigger, Auto sweep mode, 0 V level, channel 1
source, DC coupling, rising edge slope, 60 ns holdoff time.
Display
Vectors on, 33% grid intensity, infinite persistence off.
Other
Acquire mode normal, Run/Stop to Run, cursors and
measurements off.
Labels
All custom labels that you have created in the Label
Library are preserved (not erased), but all channel labels will be
set to their original names.
Summary of Contents for InfiniiVision 6000 Series
Page 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Page 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Page 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Page 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Page 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Page 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Page 362: ...362 6000 Series Oscilloscope User s Guide 9 ...