Triggering the Oscilloscope
4
6000 Series Oscilloscope User’s Guide
201
8
Press the
Bus Config
softkey and set the following parameters.
a #Bits
: Set the number of bits in the UART/RS232 words to
match your target system (selectable from 5-9 bits).
b Parity
: Choose odd, even, or none, based on your target
system.
c Baud
: Select the baud rate to match the signal in your
target system.
d Polarity
: Select idle low or idle high to match your target
system’s state when at idle. For RS232 select idle low.
e Bit Order
: Select whether the most significant bit (MSB) or
the least significant bit (LSB) is presented after the start
bit in the signal from your target system. For RS232 select
LSB.
9
Press the up arrow key to return to the previous menu.
10
Press the
Trigger Setup
softkey and set up the desired trigger
condition.
•
Rx Start Bit:
The oscilloscope triggers when a start bit
occurs on Rx.
•
Rx Stop Bit:
Triggers when a stop bit occurs on Rx. The
trigger will occur on the first stop bit. This is done
automatically whether the target system uses 1, 1.5, or 2
stop bits. You do not need to specify the number of stop
bits used by the target system.
•
Rx Data:
Triggers on a data byte that you specify. For use
when the target system data words are from 5 to 8 bits in
length (no 9th (alert) bit).
•
Rx 1:Data:
For use when the target system data words are 9
bits in length including the alert bit (the 9th bit). Triggers
only when the 9th (alert) bit is 1. The specified data byte
applies to the least significant 8 bits (excludes the 9th
(alert) bit).
•
Rx 0:Data:
For use when the target system data words are 9
bits in length including the alert bit (the 9th bit). Triggers
only when the 9th (alert) bit is 0. The specified data byte
applies to the least significant 8 bits (excludes the 9th
(alert) bit).
Summary of Contents for InfiniiVision 6000 Series
Page 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Page 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Page 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Page 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Page 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Page 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Page 362: ...362 6000 Series Oscilloscope User s Guide 9 ...