226
6000 Series Oscilloscope User’s Guide
5
Making Measurements
•
Center
— sets the FFT spectrum frequency represented at
the center vertical grid line of the display. It is possible to
set the Center to values below half the span or above the
maximum available frequency, in which case the displayed
spectrum will not take up the whole screen. Press the
Center
softkey, then turn the Entry knob to set the desired
center frequency of the display.
•
Preset
— sets the frequency Span and Center to values that
will cause the entire available spectrum to be displayed.
The maximum available frequency is half the effective FFT
sample rate, which is a function of the time per division
setting. The current FFT sample rate is displayed above
the softkeys.
2
Press the More FFT softkey to display additional FFT
settings.
•
Window
— selects a window to apply to your FFT input
signal:
•
Hanning
— window for making accurate frequency
measurements or for resolving two frequencies that are
close together.
•
Flat Top
— window for making accurate amplitude
measurements of frequency peaks.
•
Rectangular
— good frequency resolution and amplitude
accuracy, but use only where there will be no leakage
effects. Use on self-windowing waveforms such as
pseudo-random noise, impulses, sine bursts, and
decaying sinusoids.
FFT Scale
FFT Offset
Return to
previous menu
Window
Summary of Contents for InfiniiVision 6000 Series
Page 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Page 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Page 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Page 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Page 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Page 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Page 362: ...362 6000 Series Oscilloscope User s Guide 9 ...