Saving and Printing Data
7
6000 Series Oscilloscope User’s Guide
325
•
ASCII XY data file
- This creates a file of comma-separated
variables for each displayed channel. If the oscilloscope
acquisition is stopped, then data records greater than 1000
points can be written. This format is also suitable for
spreadsheets.
•
BIN data file
- This creates a binary file with a header, and
data in the form of time and voltage pairs. This file is much
smaller than the ASCII XY data file. If the oscilloscope
acquisition is stopped, then data records greater than 1000
points can be written.
•
ALB data file
- This creates a file in an Agilent proprietary
format which can be read by Agilent logic analyzers.
Choosing save settings
Press
Save/Recall
→
Save
→
Settings
to change the following
options.
These settings are not applicable when the format
“Trace&Setup” is selected.
Factors
Select
Factors
if you want the oscilloscope scale factors to be
included. Oscilloscope scale factors include vertical, horizontal,
trigger, acquisition, math, and display settings. The scale factors
will be sent to a separate file with a TXT extension.
Invert Graticule Colors
This setting is applicable when you have chosen BMP or PNG
using the
Format
softkey.
Summary of Contents for InfiniiVision 6000 Series
Page 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Page 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Page 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Page 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Page 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Page 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Page 362: ...362 6000 Series Oscilloscope User s Guide 9 ...