6000 Series Oscilloscope User’s Guide
1
Getting Started
To verify basic oscilloscope operation
If you have a 6000L Series oscilloscope, you will need to start a
Web control session, as described on
page 44
.
1
Press the
Save/Recall
key on the front panel, then press the
Default Setup
softkey. (The softkeys are located directly below
the display on the front panel.) The oscilloscope is now
configured to its default settings.
2
Connect an oscilloscope probe from channel 1 to the
Probe
Comp
signal terminal on the front panel.
3
Connect the probe’s ground lead to the ground terminal that
is next to the
Probe Comp
terminal.
4
Press
AutoScale
.
WA R N I N G
Do not negate the protective action of the ground connection to the
oscilloscope. The oscilloscope must remain grounded through its
power cord. Defeating the ground creates an electric shock hazard.
C A U T I O N
Maximum input voltage for analog inputs
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 100 Vrms, 400 Vpk
with 10073C or 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
Summary of Contents for InfiniiVision 6000 Series
Page 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Page 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Page 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Page 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Page 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Page 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Page 362: ...362 6000 Series Oscilloscope User s Guide 9 ...