:ACTivity
Command Syntax
:ACTivity
The :ACTivity command clears the cumulative edge variables for the next
activity query.
Query Syntax
:ACTivity?
The :ACTivity? query returns whether there has been activity (edges) on
the digital channels since the last query, and returns the current logic
levels.
Return Format
<edges>,<levels><NL>
<edges> ::= presence of edges (16-bit integer in NR1 format).
<levels> ::= logical highs or lows (16-bit integer in NR1 format).
bit 0 ::= DIGital 0
bit 15 ::= DIGital 15
See Also
•
"Introduction to Root (:) Commands"
•
•
N O T E
Because the :ACTivity? query returns edge activity since the last :ACTivity? query, you must
send this query twice before the edge activity result is valid.
N O T E
A bit = 0 (zero) in the <edges> result indicates that no edges were detected on that channel
(across the specified threshold voltage) since the last query.
A bit = 1 (one) in the <edges> result indicates that edges have been detected on that
channel (across the specified threshold voltage) since the last query.
(The threshold voltage must be set appropriately for the logic levels of the signals being
probed.)
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...