Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
161
:MTERegister[:EVENt] (Mask Test Event Event Register)
Query Syntax
:MTERegister[:EVENt]?
The :MTERegister[:EVENt]? query returns the integer value contained in
the Mask Test Event Event Register and clears the register.
Return Format
<value><NL>
<value> ::= integer in NR1 format.
See Also
•
"Introduction to Root (:) Commands"
•
•
To MTE bit in
Operation Status
Condition Register
+
OR
:MTEenable
:MTEenable?
Mask Test Event Enable (Mask) Register
1
0
2
3
4
5
6
7
9
8
10
11
12
13
14
15
:MTERegister[:EVENt]?
Mask Test Event Event Register
Fail
Com-
plete
Auto
Mask
Started
Table 50 Mask Test Event Event Register
Bit
Name
Description
When Set (1 = High = True), Indicates:
15-11
---
---
(Not used.)
10
Auto
Mask
Auto Mask Created
Auto mask creation completed.
9
---
---
(Not used.)
8
Started
Mask Testing Started
Mask testing started.
7-2
---
---
(Not used.)
1
Fail
Mask Test Fail
The mask test failed.
0
Comple
te
Mask Test Complete
The mask test is complete.
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...