Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
163
:OPEE (Operation Status Enable Register)
Command Syntax
:OPEE <mask>
<mask> ::= 16-bit integer
The :OPEE command sets a mask in the Operation Status Enable register.
Set any of the following bits to "1" to enable bit 7 in the Status Byte
Register and potentially cause an SRQ (Service Request interrupt to be
generated.
Wait
Trig
Run
Arm
Reg
AER?
Run bit set if oscilloscope not stopped
:OPEReration:CONDition?
Operation Status Condition Register
:OPEE
:OPEE?
Operation Status Enable
(Mask) Register
From Overload
Event Registers
To OPER bit in
Status Byte
Register
OVLR
+
OR
1
0
2
3
4
5
6
7
9
8
10
11
12
13
14
15
Wait
Trig
Run
:OPERation[:EVENt]?
Operation Status Event Register
OVLR
HWE
11
5
3
From Hardware
Event Registers
HWE
12
From Mask Test
Event Registers
MTE
MTE
9
Table 51 Operation Status Enable Register (OPEE)
Bit
Name
Description
When Set (1 = High = True), Enables:
15-13
---
---
(Not used.)
12
HWE
Hardware Event
Event when hardware event occurs.
11
OVLR
Overload
Event when 50
Ω
input overload occurs.
10
---
---
(Not used.)
9
MTE
Mask Test Event
Event when mask test event occurs.
8-6
---
---
(Not used.)
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...