What's New
1
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
23
Discontinued
Commands
Most of the following commands have beed discontinued because they are
not supported by the new N5432C FlexRay triggering and decode option
(Option FLX) which replaces previous FlexRay triggering and serial decode
options.
:TRIGger:FLEXray:EVENt:TYPE (see
The BSS (Byte Start Sequence) has been added
and the FSS (Frame Start Sequence) has been
removed.
:TRIGger:FLEXray:TRIGger (see
The TIME trigger type is no longer supported.
:TRIGger:MODE (see
You can now select the M1553 trigger mode.
In the RAW or MAXimum waveform points
modes, you can now specify 4,000,000 or
8,000,000 points in place of the previous
5,000,000 option.
Command syntax is the same, but the NORMal
mode returns:
•
The measurement record when
:SYSTem:PRECision is OFF.
•
The precision analysis record when
:SYSTem:PRECision is ON.
Command
Differences
Discontinued Command
Current Command Equivalent
Comments
:DISPlay:FREeze
none
:SBUS:BUSDoctor:ADDRess
none
The VPT1000 (BusDoctor)
vehicle protocol tester module
is not used with the new FLX
option.
:SBUS:BUSDoctor:BAUDrate
none
The VPT1000 (BusDoctor)
vehicle protocol tester module
is not used with the new FLX
option. You now specify the
baud rate using the
:TRIGger:FLEXray:BAUDrate
(see
:SBUS:BUSDoctor:CHANnel
none
The VPT1000 (BusDoctor)
vehicle protocol tester module
is not used with the new FLX
option. You now specify bus A
or B using the
:TRIGger:FLEXray:CHANnel
(see
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...