Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
5
Commands by Subsystem
:MEASure:VBASe
Command Syntax
:MEASure:VBASe [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VBASe command installs a screen measurement and starts a
waveform base value measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax
:MEASure:VBASe? [<source>]
The :MEASure:VBASe? query returns the vertical value at the base of the
waveform. The base value of a pulse is normally not the same as the
minimum value.
Return Format
<base_voltage><NL>
<base_voltage> ::= value at the base of the selected waveform in
NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
•
•
N O T E
This command is not available if the source is FFT (Fast Fourier Transform).