Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
365
:MTESt:AMASk:UNITs
Command Syntax
:MTESt:AMASk:UNITs <units>
<units> ::= {CURRent | DIVisions}
The :MTESt:AMASk:UNITs command alters the way the mask test
subsystem interprets the tolerance parameters for automasking as defined
by :MTESt:AMASk:XDELta and :MTESt:AMASk:YDELta commands.
•
CURRent — the mask test subsystem uses the units as set by the
:CHANnel<n>:UNITs command, usually time for
Δ
X and voltage for
Δ
Y.
•
DIVisions — the mask test subsystem uses the graticule as the
measurement system, so tolerance settings are specified as parts of a
screen division. The mask test subsystem maintains separate XDELta
and YDELta settings for CURRent and DIVisions. Thus, XDELta and
YDELta are not converted to new values when the UNITs setting is
changed.
Query Syntax
:MTESt:AMASk:UNITs?
The :MTESt:AMASk:UNITs query returns the current measurement units
setting for the mask test automask feature.
Return Format
<units><NL>
<units> ::= {CURR | DIV}
See Also
•
"Introduction to :MTESt Commands"
•
•
•
•
•
Example Code
•
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...