:MTESt:SCALe:BIND
Command Syntax
:MTESt:SCALe:BIND <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:SCALe:BIND command enables or disables Bind 1 & 0 Levels
(Bind - 1 & 0 Levels for inverted masks) control:
•
ON —
If the Bind 1 & 0 Levels control is enabled, the 1 Level and the 0 Level
controls track each other. Adjusting either the 1 Level or the 0 Level
control shifts the position of the mask up or down without changing its
size.
If the Bind - 1 & 0 Levels control is enabled, the - 1 Level and the 0
Level controls track each other. Adjusting either the - 1 Level or the 0
Level control shifts the position of the mask up or down without
changing its size.
•
OFF —
If the Bind 1 & 0 Levels control is disabled, adjusting either the 1 Level
or the 0 Level control changes the vertical height of the mask.
If the Bind - 1 & 0 Levels control is disabled, adjusting either the - 1
Level or the 0 Level control changes the vertical height of the mask.
Query Syntax
:MTESt:SCALe:BIND?
The :MTESt:SCALe:BIND? query returns the value of the Bind 1&0 control
(Bind - 1&0 for inverted masks).
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also
•
"Introduction to :MTESt Commands"
•
•
•
•
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...