:TRIGger:FLEXray:EVENt:TYPE
Command Syntax
:TRIGger:FLEXray:EVENt:TYPE <event>
<event> ::= {WAKeup | TSS | {FES | DTS} | BSS}
Selects the FlexRay event to trigger on. The event setting is only valid
when the FlexRay trigger mode is set to EVENt.
•
WAKeup — triggers on Wake- Up event.
•
TSS — triggers on Transmission Start Sequence event.
•
FES — triggers on Frame End Sequence event.
•
DTS — triggers on Dynamic Trailing Sequence event.
•
BSS — triggers on Byte Start Sequence event.
Query Syntax
:TRIGger:FLEXray:EVENt:TYPE?
The :TRIGger:FLEXray:EVENt:TYPE? query returns the currently selected
FLEXray event.
Return Format
<event><NL>
<event> ::= {WAK | TSS | {FES | DTS} | BSS}
See Also
•
"Introduction to :TRIGger Commands"
•
•
•
N O T E
FES and DTS are equivalent.
N O T E
This command is only valid when the FLEXray triggering and serial decode option (Option
FLX) has been licensed.