Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
545
:TRIGger:IIC:PATTern:ADDRess
Command Syntax
:TRIGger:IIC:PATTern:ADDRess <value>
<value> ::= integer or <string>
<string> ::= "0xnn" where n ::= {0,..,9 | A,..,F}
The :TRIGger:IIC:PATTern:ADDRess command sets the address for IIC
data.The address can range from 0x00 to 0x7F (7- bit) or 0x3FF (10- bit)
hexadecimal. Use the don't care address (- 1 or 0xFFFFFFFF) to ignore the
address value.
Query Syntax
:TRIGger:IIC:PATTern:ADDRess?
The :TRIGger:IIC:PATTern:ADDRess? query returns the current address for
IIC data.
Return Format
<value><NL>
<value> ::= integer
See Also
•
"Introduction to :TRIGger Commands"
•
•
•