576
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
5
Commands by Subsystem
:TRIGger:SEQuence:FIND
Command Syntax
:TRIGger:SEQuence:FIND <value>
<value> ::= {PATTern1,ENTered | PATTern1,EXITed | EDGE1
| PATTern1,AND,EDGE1}
The :TRIGger:SEQuence:FIND command specifies the find stage of a
sequence trigger. This command accepts three program data parameters;
you can use NONE to fill out the parameter list (for
example,"EDGE1,NONE,NONE").
PATTern1 is specified with the":TRIGger:SEQuence:PATTern command.
EDGE1 is specified with the :TRIGger:SEQuence:EDGE command.
Query Syntax
:TRIGger:SEQuence:FIND?
The :TRIGger:SEQuence:FIND? query returns the find stage specification
for a sequence trigger. NONE is returned for unused parameters.
Return Format
<find_value><NL>
<find_value> ::= {PATT1,ENT,NONE | PATT1,EXIT,NONE | EDGE1,NONE,NONE
| PATT1,AND,EDGE1}
See Also
•
"Introduction to :TRIGger Commands"
•
•
•
•
Summary of Contents for InfiniiVision 7000A Series