:CHANnel:ACTivity
Command Syntax
:CHANnel:ACTivity
The :CHANnel:ACTivity command clears the cumulative edge variables for
the next activity query.
Query Syntax
:CHANnel:ACTivity?
The :CHANnel:ACTivity? query returns the active edges since the last clear,
and returns the current logic levels.
Return Format
<edges>,<levels><NL>
<edges> ::= presence of edges (32-bit integer in NR1 format).
<levels> ::= logical highs or lows (32-bit integer in NR1 format).
A bit equal to one indicates that edges have been detected at the specified
threshold since the last clear on that channel.
N O T E
The :CHANnel:ACTivity command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :ACTivity command (see
) instead.
N O T E
A bit equal to zero indicates that no edges were detected at the specified threshold since
the last clear on that channel. Edges may have occurred that were not detected because of
the threshold setting.
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...