Commands Quick Reference
Table 11 :FUNCtion Commands Summary
Command
Query
Options and Query Returns
:FUNCtion:CENTer
<frequency> (see
:FUNCtion:CENTer?
)
<frequency> ::= the current
center frequency in NR3 format.
The range of legal values is from
0 Hz to 25 GHz.
:FUNCtion:DISPlay {{0
| OFF} | {1 | ON}}
(see
:FUNCtion:DISPlay?
)
{0 | 1}
:FUNCtion:GOFT:OPERat
ion <operation> (see
:FUNCtion:GOFT:OPERat
)
<operation> ::= {ADD | SUBTract |
MULTiply}
:FUNCtion:GOFT:SOURce
1 <source> (see
:FUNCtion:GOFT:SOURce
)
<source> ::= CHANnel<n>
<n> ::= {1 | 2 | 3 | 4} for 4ch
models
<n> ::= {1 | 2} for 2ch models
:FUNCtion:GOFT:SOURce
2 <source> (see
:FUNCtion:GOFT:SOURce
)
<source> ::= CHANnel<n>
<n> ::= {{1 | 2} | {3 | 4}} for
4ch models, depending on SOURce1
selection
<n> ::= {1 | 2} for 2ch models
:FUNCtion:OFFSet
<offset> (see
:FUNCtion:OFFSet?
)
<offset> ::= the value at center
screen in NR3 format.
The range of legal values is
+/-10 times the current
sensitivity of the selected
function.
:FUNCtion:OPERation
<operation> (see
:FUNCtion:OPERation?
)
<operation> ::= {ADD | SUBTract |
MULTiply | INTegrate |
DIFFerentiate | FFT | SQRT}
:FUNCtion:RANGe
<range> (see
:FUNCtion:RANGe? (see
)
<range> ::= the full-scale
vertical axis value in NR3
format.
The range for ADD, SUBT, MULT is
8E-6 to 800E+3. The range for the
INTegrate function is 8E-9 to
400E+3.
The range for the DIFFerentiate
function is 80E-3 to 8.0E12
(depends on current sweep speed).
The range for the FFT function is
8 to 800 dBV.
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...