Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
7
Obsolete and Discontinued Commands
:MEASure:TSTOp
Command Syntax
:MEASure:TSTOp <value> [suffix]
<value> ::= time at the stop marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTOp command moves the stop marker (X2 cursor) to the
specified time with respect to the trigger time.
Query Syntax
:MEASure:TSTOp?
The :MEASure:TSTOp? query returns the time at the stop marker (X2
cursor).
Return Format
<value><NL>
<value> ::= time at the stop marker in NR3 format
See Also
•
"Introduction to :MARKer Commands"
•
"Introduction to :MEASure Commands"
•
•
•
•
•
N O T E
The short form of this command, TSTO, does not follow the defined Long Form to Short
Form Truncation Rules (see
). The normal short form "TST" would be the same for
both TSTArt and TSTOp, so sending TST for the TSTOp command produces an error.
N O T E
The :MEASure:TSTOp command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:X2Position command (see
) instead.
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...