Status Reporting
Status Reporting Data Structures
The following figure shows how the status register bits are masked and
logically OR'ed to generate service requests (SRQ) on particular events.
9
8
10
11
12
13
14
15
+
OR
1
0
2
3
4
5
6
7
:OVLR?
Overload Event Register
:OVL
:OVL?
Overload Event Enable (Mask) Register
Chan2
OVL
Chan1
OVL
Chan3
OVL
Chan4
OVL
Ext Trig
OVL
Chan2
Fault
Chan1
Fault
Chan3
Fault
Chan4
Fault
Ext Trig
Fault
Bat
ON
:HWERegister:CONDition?
Hardware Event Condition Register
:HWEenable
:HWEenable?
Hardware Event Enable (MASK) Register
1
0
2
3
4
5
6
7
9
8
10
11
12
13
14
15
:HWERegister[:EVENt]?
Hardware Event Event Register
PLL
Locked
0
Bat
ON
+
OR
+
OR
:MTEenable
:MTEenable?
Mask Test Event Enable (MASK) Register
1
0
2
3
4
5
6
7
9
8
10
11
12
13
14
15
:MTERegister[:EVENt]?
Mask Test Event Event Register
Com-
plete
Fail
MTE
To bits in Operation Status Condition Register:
OVLR
HWE
Auto
Mask
Started
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...